CPI Indico 100 Series Service And Installation Manual page 232

Hide thumbs Also See for Indico 100 Series:

Advertisement

5
Troubleshooting
5.2.3
Error Messages (Cont)
ERROR CODE
E008
E009
E011
E012
E013
E014
E015
Page 5-4
MESSAGE
KV/MA FAULT
PS NOT READY
HIGH MA FAULT
LOW MA FAULT
MANUAL TERMIN
AEC BUT ERROR
AEC BU MAS ERR
Rev. B
Millenia / Indico 100 Series Service Manual
PROBLEM
Power supply has detected
a fault in the kV or mA
output during an exposure
and immediately terminated
the exposure. This may be
caused by arcing in the X-
ray tube, arcing of the HV
cables, or HT tank.
Power supply is not ready to
make an exposure.
Generator CPU detected
mA greater than allowed
tolerance.
Generator CPU detected
mA less than allowed
tolerance.
Operator released exposure
switch during exposure.
AEC exposure exceeded
allowed back up time.
AEC exposure exceeded
allowed back up mAs.
CPI Canada Inc
ACTION
1. If arcing of the X-
ray tube is
suspected, check
condition of tube.
The X-ray tube
may be damaged
or simply require
"seasoning".
Refer to chapter 6
for tube
seasoning
procedure.
2. If failure of HT
tank is suspected,
contact product
support.
Retry exposure.
Recalibrate X-ray
tube.
Recalibrate X-ray
tube.
1. Re-take exposure
if necessary.
2. Check for faulty
switch contacts or
wiring.
1. Check exposure
technique
settings.
2. Check that correct
AEC chamber is
energized.
1. Check exposure
technique
settings.
2. Check that correct
AEC chamber is
energized.
Ch # 740904-09