CPI Indico 100 Series Service And Installation Manual page 234

Hide thumbs Also See for Indico 100 Series:

Advertisement

5
Troubleshooting
5.2.3
Error Messages (Cont)
ERROR CODE
E032
E033
E034
E035
E036
E037
E038
E039
E040
E041
E042
E043
E044
E045
E046
E047
Page 5-6
MESSAGE
CONSOLE COMM ERR
GEN BATTERY LOW
+12VDC ERROR
-12VDC ERROR
+15VDC ERROR
-15VDC ERROR
CAL DATA ERROR
AEC DATA ERROR
FLUORO DATA ERROR
REC DATA ERROR
TUBE DATA ERR
KV ERROR
COMM ERROR
NOT SUPPORTED
MODE INHIBITED
FL TIMER LIMIT
Rev. B
Millenia / Indico 100 Series Service Manual
PROBLEM
Generator has detected
error in communication to
console.
Generator detects lithium
battery voltage is low.
+12VDC rail is out of
tolerance.
-12VDC rail is out of
tolerance.
+15VDC rail is out of
tolerance.
-15VDC rail is out of
tolerance.
Generator detects corrupt
calibration data.
Generator detects corrupt
AEC data.
Generator detects corrupt
Fluoro data.
Generator detects corrupt
receptor data.
Generator detects corrupt
tube data.
KV detected in non x-ray
state.
Received communication
message not valid and
ignored.
Received message valid,
but not supported by this
system.
Received message valid,
but not allowed during
present state.
Fluoro Timer has exceeded
time limit.
CPI Canada Inc
ACTION
1. Check console
cable for damage
and proper
connection.
2. Turn power off and
then on to reset
generator.
Replace lithium
battery.
Check +12VDC rail.
Check -12VDC rail.
Check +15VDC rail.
Check -15VDC rail.
Re-calibrate X-ray
tube(s).
Reprogram AEC data
or set factory
defaults.
Reprogram fluoro
data or set factory
defaults.
Reprogram receptor
data or set factory
defaults.
Reprogram tube data
or set factory
defaults.
Switch OFF
generator. Prevent
further use of
generator. Call
product support.
Reset error.
Reset error.
Reset error.
Reset Fluoro timer.
Ch # 740904-09