X-Ray Generator Update/Service Record; Maintenance Schedule - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

6
Regular Maintenance
6.2.0

X-RAY GENERATOR UPDATE/SERVICE RECORD

The X-ray generator update/service record is stored in the upper cabinet of the generator (Millenia) or on the
back of the panel which accesses the control boards in the lower cabinet (Indico 100). The installation date and
location should be recorded on this form at the time of the original site installation.
Service and repairs must be recorded in the update/service record. The record should be as thorough
as possible, detailing the scope and type of work that was performed (all service and a record of all
replacement parts that were installed). Additionally, the person performing the work should date and sign the
record.
This information will be invaluable in the future for traceability and to ensure continued compatibility of
the generator.
6.3.0

MAINTENANCE SCHEDULE

WARNING:
MAINTENANCE IS TO BE PERFORMED ONLY BY COMPETENT, TRAINED PERSONNEL WHO
ARE FAMILIAR WITH THE POTENTIAL HAZARDS ASSOCIATED WITH THIS EQUIPMENT.
NOTE:
MAINTENANCE SCHEDULE FREQUENCY MAY BE DETERMINED BY CERTAIN REGULATORY
REQUIREMENTS OF THE COUNTRY OR STATE IN WHICH THE INSTALLATION IS LOCATED.
ALWAYS CHECK THE LOCAL CODES AND REGULATIONS WHEN DETERMINING A
MAINTENANCE SCHEDULE.
WARNING: ALWAYS SWITCH OFF MAINS POWER TO THE GENERATOR AND WAIT A
MINIMUM OF 5 MINUTES FOR CAPACITORS TO DISCHARGE BEFORE
BEGINNING ANY PREVENTATIVE MAINTENANCE, INCLUDING CLEANING.
WARNING: OBSERVE ESD PRECAUTIONS. KEEP ALL STATIC - SENSITIVE COMPONENTS
AND CIRCUIT BOARDS IN THEIR STATIC - SHIELDING PACKAGING UNTIL
READY TO INSTALL. ENSURE THAT YOU ARE GROUNDED AT ALL TIMES
WHEN HANDLING STATIC - SENSITIVE COMPONENTS AND CIRCUIT BOARDS.
Page 6-2
Rev. D
Millenia / Indico 100 Series Service Manual
CPI Canada Inc
Ch # 740904-10