CPI Indico 100 Series Service And Installation Manual page 200

Hide thumbs Also See for Indico 100 Series:

Advertisement

3E
ABS Calibration
3E.3.4 Dose1/Dose2 Calibration (Cont)
7.
8.
9.
Page 3E-22
Reset the original NOMINAL DOSE value
by entering the value recorded in step 1 of
this table into the NOMINAL DOSE
location using the + or - buttons.
Verify each dose (nominal, dose 1, and
dose 2) by initiating fluoro operation and
measuring the I.I. input dose.
Press <<, then EXIT to exit out of
FLUORO SETUP mode. Press EXIT, then
EXIT SETUP to return to the normal
operating mode.
This completes the fluoro calibration.
Rev. D
Millenia / Indico 100 Series Service Manual
CPI Canada Inc.
Ch # 740904-07