Download Print this page

CPI Indico 100 Series Service And Installation Manual page 260

Hide thumbs Also See for Indico 100 Series:

Advertisement

8
Spares
NOTE (Cont):
5.
Three different inverter board part numbers are used in Indico 100 generators. For 230 VAC 1 phase
and 480 VAC 3 phase generators use part number 732813-00, for 400 VAC 3 phase 30 / 50 / 80 kW
generators use part number 732813-04, for 230 VAC 1 phase 37.5 kW and for 400 VAC 3 phase 37.5 /
65 kW generators use part number 732813-05 (see note 17 for the definition of mains input voltage). 30
and 37.5 kW generators require one inverter board, 50 and 65 kW generators require two inverter
boards, and 80 kW generators require three inverter boards. Spares should be stocked accordingly.
6a.
Two part numbers of power input boards are used for spares. For 1 phase 230 VAC generators use part
number 733798-01, for 3 phase 400 or 480 VAC generators use part number 732161-01.
6b.
This fuse is used on all rad two tube generators, and on all R & F one tube and two tube generators.
Spares should be stocked accordingly.
7.
Five different resonant boards are used in Indico 100 generators. Refer to the table below for the correct
part number for your generator.
O/P POWER (kW) & MAINS VOLTS
30 kW (230 VAC 1φ, 400 / 480 VAC 3φ)
37.5 / 40 kW (230 VAC 1φ, 400 VAC 3φ)
80 kW (400 / 480 VAC 3φ)
8.
The AEC board for your generator was selected to be compatible with specific AEC device(s). To
maintain full compatibility, the original part number must be ordered as a replacement. Refer to chapter
9, section 9.2.0 for the part number of the original AEC board shipped in the generator for which this
manual was prepared.
9.
The console display board is supplied pre-mounted to the console front panel switch assembly. To
determine the required spares part number for this assembly, refer to the table below. CONSOLE
DISPLAY BOARD ONLY USED WITH THE 23 X 56 CM CONSOLE.
SWITCH/DISPLAY ASSY, R&F, ENGLISH
SWITCH/DISPLAY ASSY, R&F, GERMAN
SWITCH/DISPLAY ASSY, RAD, ENGLISH
SWITCH/DISPLAY ASSY, RAD, GERMAN
10.
Remote fluoro control is an option, spares should be stocked accordingly.
11.
These items only used on generators fitted with low speed starter, spares should be stocked accordingly.
Page 8-4
50 kW (400 VAC 3φ)
50 kW (480 VAC 3φ)
65 kW (400 VAC 3φ)
80 kW (400 VAC 3φ)
DESCRIPTION
Rev. K
RESONANT BOARD
RAD POWER SUPPLIES
732808-00
732808-05
732808-01
732808-02
732808-06
732808-03
SWITCH/DISPLAY ASSY PART NO
Indico 100 Series Service Manual
CPI Canada Inc
RESONANT BOARD
R&F POWER SUPPLIES
732964-02
732964-00
732964-03
732964-01
734429-00
734429-01
734430-00
734430-01
Ch # 740895-07

Advertisement

loading