Download Print this page

Dip Switch Settings; Initial Run-Up; Initial Voltage Measurements - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc.

2.17.0 DIP SWITCH SETTINGS

Before continuing, verify the DIP switch settings on the console CPU board and generator CPU board.
These switches have been factory set but may have been readjusted, particularly if this generator is a re-
install.
CONSOLE CPU BOARD (23 X 56 cm console) / CONSOLE BOARD (31 X 42 cm console)
Verify the settings on SW1, the switches should all be set to OFF. Setting SW1-8 ON at power up
restores factory default settings for the console. SW1-1 to SW1-7 should remain in the OFF position
at all times.
GENERATOR CPU BOARD:
Verify the settings on SW1. Refer to the table below for the proper settings for this switch.
GENERATOR POWER
30 kW
37.5 kW
50 kW
65 kW
80 kW
OFF > 2 FILAMENT BOARDS
SW1-4
SW1-5
OFF > 150 kV MAXIMUM
OFF > DUAL SPEED STARTER
SW1-6
OFF > 2 TUBE GENERATOR
SW1-7
SW1-8
OFF > SET FACTORY DEFAULTS

2.18.0 INITIAL RUN-UP

This section describes the procedure for initial power-on of the generator after it has first been installed.

2.18.1 Initial Voltage Measurements

1.
Verify that the mains voltage and current capacity is correct for the generator installation. Refer to
the product ID label on the generator cabinet and chapter 1C of this manual.
2.
Temporarily remove the safety cover over the main input fuses in the generator.
3.
If the mains supply is compatible with the generator, switch on the main breaker and/or
disconnect switch and check for the following voltages:
NOTE:
DO NOT SWITCH ON THE GENERATOR AT THIS TIME (ONLY THE AC MAINS TO THE
GENERATOR IS TO BE SWITCHED ON AT THIS TIME).
Indico 100 Series Service Manual
MAXIMUM mA
320 mA
400 mA
630 mA
800 mA
1000 mA
Ch # 740895-04
SW1 -3
SW1-2
ON
ON
ON
OFF
OFF
ON
OFF
ON
OFF
OFF
ON > I FILAMENT BOARD
ON > 125 kV MAXIMUM
ON > LOW SPEED STARTER
ON > 1 TUBE GENERATOR
ON > DEFAULTS DISABLED
Rev. H
2
Installation
SW1-1
OFF
ON
ON
OFF
ON
Page 2-31

Advertisement

loading