X-Ray Tube Stator & Thermal Switch Connections
Refer to figure 2-6 for the X-ray tube stator and thermal switch connections.
1.
Route the X-ray tube stator cable(s) through the lower wiring access panel panel on the rear of
the generator cabinet, then route the cables towards the stator terminal blocks as shown in the
figure below.
FOR UNITS WITH A LOW SPEED STARTER, SHIELDED STATOR CABLES ARE
RECOMMENDED. FOR UNITS WITH A DUAL SPEED STARTER, SHIELDED STATOR
CABLES MUST BE USED.
THE CHASSIS GROUND STUD(S) SHOWN IN FIGURE 2-6.
2.
Connect the wires to the appropriate terminal as shown. The tube thermal switch will normally be
connected to the THERMAL SWITCH connections on the stator terminal block, but may optionally
be connected to the room interface board (refer to chapter 3B).
3.
Ensure that all terminal connections are tight, then dress and secure the cables.
NOTE:
ONE TUBE ONLY GENERATORS WILL HAVE 1 STATOR TERMINAL BLOCK FITTED. THE
LOWER TERMINAL BLOCK IN FIGURE 2-6 IS ONLY FITTED ON TWO TUBE GENERATORS.
Page 2-12
THE SHIELD FOR THE STATOR CABLE(S) MUST BE PROPERLY GROUNDED AT
Figure 2-6: Stator connections to generator
Rev. H
Indico 100 Series Service Manual
CPI Canada Inc.
Ch # 740895-04