Download Print this page

CPI Indico 100 Series Service And Installation Manual page 231

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc
5.2.2
Limit Messages (Cont)
GEN DUTY WARNING
5.2.3
Error Messages
These messages indicate that an error has occurred. The errors are logged in the error log, previous errors
should be reviewed by service personnel before taking further action.
ERROR CODE
E001
E003
E004
E005
E006
E007
Millenia / Indico 100 Series Service Manual Ch # 740904-09
The generator has exceeded its duty cycle
warning limit.
MESSAGE
GEN EPROM ERR
GEN NVRAM ERR
GEN RTC ERROR
PS CONTACT ERR
ROTOR FAULT
FILAMENT FAULT
PROBLEM
Generator CPU EPROM has
been corrupted.
Generator CPU NVRAM
data has been corrupted.
Generator CPU real time
clock is not functioning.
Main contactor in HF power
supply did not energize.
1. Rotor starter may have
detected a current fault in
the stator.
2. Power supply was not
ready to start rotor.
Power supply has detected
filament current <2 amps.
Rev. B
Troubleshooting
Re-evaluate technique
factors. Allow
generator to cool if
possible. If exposures
are continued, serious
generator damage
may result due to
overheating.
ACTION
Call product support
for new generator
CPU EPROM.
Re-initialize
generator CPU
NVRAM using
generator factory
defaults.
Reset time and date.
Call product support.
Power unit off and
retry rotor start.
1. Check for open
filament in X-ray
tube.
2. Check for poor
connections in the
cathode cable.
3. Check fuses on
filament board(s).
Page 5-3