Reproducibility, Linearity, & HVL Testing
4.0
LINEARITY (Cont)
Record the mAs and mR values taken from tables 3, 4, 6, and 7 below.
Table 3
Table 4
Table 6
Table 7
X
=
3
X
=
6
L =
In the numerator of the above equation, use the absolute value of X
(disregard the minus sign).
X
=
4
X
=
7
L =
In the numerator of the above equation, use the absolute value of X
(disregard the minus sign).
Page 8
mAs
= _____
3
mAs
= _____
4
mAs
= _____
6
mAs
= _____
7
mR
3
= ________
mAs
3
mR
6
= ________
mAs
6
X
- X
3
6
= ________ (not to exceed 0.095)
X
+ X
3
6
Table 8: Linearity
mR
4
= ________
mAs
4
mR
7
= ________
mAs
7
X
- X
4
7
= ________ (not to exceed 0.095)
X
+ X
4
7
Table 9: Linearity
Rev. A
X-Ray Generator Service Manual Supplement
mR
= _____
3
mR
= _____
4
mR
= _____
6
mR
= _____
7
CPI Canada Inc
- X
3
6
- X
4
7
740917-00