Preparing for Installation
1C.2.7 80 kW Three Phase
Line Voltage
Line Frequency
Momentary Current
Standby Current
*NOTE:
THREE PHASE INDICO 100 GENERATORS ARE AVAILABLE IN 400 VAC AND 480 VAC
MODELS. 400 VAC MODELS MUST BE OPERATED FROM 400 VAC MAINS, OR MAY BE
OPERATED
TRANSFORMER. 480 VAC MODELS MUST BE OPERATED FROM 480 VAC MAINS (THESE
ARE THE NOMINAL MAINS VOLTAGES, THE ALLOWED TOLERANCES ARE AS DETAILED
IN THE PREVIOUS TABLES).
1C.2.8 Service Disconnect (All Models)
Refer to the following table for recommended service disconnect ratings.
Page 1C-4
400 VAC ± 10%, 3∅ (for 400 VAC input generators).
480 VAC ± 10%, 3∅ with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC ± 10%, 3∅ (for 480 VAC input generators).
*
50/60 Hz.
200 Amps/phase at 400 VAC.
180 Amps/phase at 480 VAC.
5 Amps.
FROM
480
VAC
MAINS
Rev. D
SEE NOTE AT THE END OF 1C.2.7
WITH
AN
Indico 100 Series Service Manual
CPI Canada Inc.
OPTIONAL
LINE
ADJUSTING
Ch # 740895-02