3C.5.3 Receptor Setup (cont)
Definitions of receptor setup programming as used in this section follow. The selections made
when programming a receptor apply to that receptor only.
TUBE
TOMO
FLUORO
SERIAL
AEC CHANNEL
INTERFACE OPTS
RECEPTOR SYM
FLUORO HANG
RAD HANG
LAST IMAGE HOLD
MEMORY
Millenia / Indico 100 Series Service Manual
Selects the tube assigned to that receptor. Selecting NONE
disables that receptor.
Enables or disables tomographic operation (NO is disabled).
Enables or disables fluoroscopic operation (NO is disabled).
Allows repeated (serial) X-ray exposures without the need to re-
prep after each exposure. Normally used with serial film
changers (NO is disabled).
Defines which AEC channel will be used by the receptor. This
must be set to a valid channel or 0 as noted below. For example
if using an AEC board with only 3 input channels (channels 1 to
3) then selecting AEC channel 4 will cause an error. Selecting 0
disables AEC operation on that receptor.
Selects pre-defined interface options:
0 = None
1 = InfiMed digital
2 = ATS ESI digital
3 = Gilardoni digital pulsed RAD
4 = Gilardoni digital HCF
5 = Not assigned at this time
Allows one of six receptor symbols [sym] to be associated with
the selected receptor. Symbol 7 is blank and may be used for
an auxiliary device.
Sets the time that the rotor will continue spinning after a fluoro
exposure has terminated.
Sets the time that the rotor will continue spinning after a rad
exposure has terminated.
Sets the time that the exposure will continue after the fluoro
footswitch has been released. This enables a frame store
device to complete the last image.
Defines the techniques that will be defaulted to when a receptor
is selected:
YES: The selected receptor will remember it's last techniques
such that those techniques are displayed when that receptor is
re-selected.
NO: The selected receptor will not remember the last
techniques used on that receptor. The techniques used will be
the same as last used on the previous receptor.
DEF: The techniques used for that receptor will be as
programmed. See receptor setup menu 4 and 5
Ch # 740904-05
Programming the Generator
Rev. B
3C
Page 3C-13