1B.7.2 Caution HV Label
This label is attached to the power input board, and to the resonant board assembly. The input power
board has line voltage components attached that will be live at all times that the main disconnect is
switched on. This includes the input fuses, main line contactor, mains rectifier and associated
components.
The resonant board assembly has components that may be energized at all times that the
generator is on. These components may retain their charge for 5 minutes after the console is switched off,
or the main disconnect is switched off.
1B.7.3 Danger HV Label
This label is attached to the primary terminals on the HT oil tank. These terminals may be energized at all
times that the generator is switched on, and for 5 minutes after the console or the main disconnect is
switched off.
1B.7.4 High Voltage Warning Label
This label is attached to the inverter board(s) and to the low speed starter assembly if fitted. The inverter
assembly is connected to the main DC bus and will have high voltage applied at all times that the
generator is switched on. This assembly will remain energized for 5 minutes after the generator is
switched off, or the main disconnect is off.
Components on the low speed starter assembly will have 240 VAC applied when the console is
switched on.
Millenia / Indico 100 Series Service Manual
Ch # 740904-01
Rev. D
1B
Safety
Page 1B-13