1C.7.0 ENVIRONMENTAL REQUIREMENTS
Listed below are the ventilation requirements for the Indico 100 series generator:
1C.8.0 CABLES SUPPLIED WITH THE INDICO 100 GENERATOR
Figure 1C-5 shows the cabling supplied with the generator:
R E M O TE FL U O R O
(O P TIO NA L)
Indico 100 Series Service Manual
•
Unrestricted air flow must be provided at the front and sides of the cabinet, as well as
underneath the unit.
•
Do not allow storage on top of the cabinet.
•
Typical heat output is 4000 BTU/hr (fluoro operation).
•
Control console heat output is negligible (150 BTU/hr).
•
The cable supplied for the console is a 15 conductor cable with a standard length of
50 ft. (15 m).
•
The cable supplied for the optional remote fluoro control is a 9 conductor cable with a
standard length of 50 ft. (15 m).
50 ' (15 M ) S U P P LIE D
G E N ER ATO R
C A BINE T
Figure 1C-5: Cabling supplied with generator
Ch # 740895-02
Preparing for Installation
50 ' (15 M ) S U P P LIE D
IN D IC O 10 0 G E N E R ATO R S
U S E O N E O F T H E T W O
C O N S O L E S T Y LE S
S H O W N IN T H IS F IG U R E
23 X 56 C M C O N S O LE
Rev. D
1C
31 X 42 C M
C O N S O LE