Programming the Generator
3C.5.3 Receptor Setup (cont)
REM TOMO BUT
SF/LF SWITCH
AEC BACKUP
AEC BACKUP MAS
AEC BACKUP MS
DEFAULTS
THE FOLLOWING SELECTIONS ARE ONLY AVAILABLE IF DEFAULTS WAS ENABLED AS
PREVIOUSLY DESCRIBED.
TECHNIQUE
FOCUS
FILM SCREEN
LEFT FIELD
CENTER FIELD
RIGHT FIELD
KV
MA
MS
DENSITY
Page 3C-14
Sets the default tomo backup time when tomo is selected via the
REMOTE TOMO SELECT input.
AUTO: Small or large focus will automatically be selected by the
generator depending on the tube current.
MAN: The operator must manually select small/large focus.
Defines the AEC backup mode to be used:
FIXED: The generator will determine the maximum AEC backup
time, not to exceed preset AEC backup mAs/ms values or
system limits. The characters AEC will be displayed in the time
window of the LED display during AEC operation.
MAS: Allows the operator to adjust the AEC backup mAs, not to
exceed preset AEC backup mAs/ms values or system limits.
The mAs value will be displayed in the time window of the LED
display during AEC operation.
MS: Allows the operator to adjust the AEC backup ms, not to
exceed preset AEC backup mAs/ms values or system limits.
The ms value will be displayed in the time window of the LED
display during AEC operation.
Sets the maximum backup mAs, to a limit of 500 mAs (600 mAs
for some generator models).
Sets the maximum back-up ms.
This selection is available only if MEMORY was set to DEF in
RECEPTOR SETUP menu 2.
Defines which technique will be defaulted to when a receptor is
selected. Options are AEC, MA, MAS.
Defines which focus will be defaulted to when a receptor is
selected. Options are LARGE or SMALL.
Defines which film screen will be defaulted to when a receptor is
selected and AEC enabled. Options are film screen 1, 2, or 3.
Selects the left field on the AEC device when AEC is selected.
As above, but for center field.
As above, but for right field.
Selects the default kV value.
Selects the default mA value.
Selects the default time value.
Selects the default density value.