CPI Indico 100 Series Service And Installation Manual page 190

Hide thumbs Also See for Indico 100 Series:

Advertisement

3E
ABS Calibration
3E.3.1 Dose Limits (Cont)
UTILITY
EXIT SETUP
TUBE SELECTION
GENERATOR LIMITS
RECEPTOR SETUP
I/O CONFIGURATION
EXIT
TUBE1
DOSE LIMITS
ABS SETUP
MIN FLUORO KV:
EXIT
TUBE1
LOOP GAIN
NOMINAL DOSE
DOSE 1:
DOSE 2:
Use these steps to set the DOSE LIMITS.
Step
1.
2.
3.
4.
5.
6.
7.
8.
Page 3E-12
* GENERATOR SETUP *
* GEN CONFIGURATION *
0%HU
FLUORO SETUP*
50
0%HU
*ABS SETUP*
60
80
100
100
<<
Set up the radiation probe as per figure 3E-5 in the position indicated Max. Patient
Input Dose. Refer to the note on page 10 re focal spot to radiation probe distance. No
absorber is required at this point in the setup.
Temporarily unplug the ABS pickup at J7 or J8 of the generator interface board.
Temporarily de-energize the I.I. power supply, or cover the I.I. with approximately 1/16"
(1.6 mm) lead.
Enter into the generator programming mode. Refer to chapter 3C, section 3C.1.1.
From the GENERATOR SETUP menu select GEN CONFIGURATION.
From the GEN CONFIGURATION menu select FLUORO SETUP.
Select ABS SETUP. Select AUTO MA/KV CURVE, then use the + or - buttons to
select auto mA/kV curve 1.
Select << to return to the FLUORO SETUP menu.
Rev. D
Millenia / Indico 100 Series Service Manual
APR EDITOR: DISABLED
GEN CONFIGURATION
DATA LINK
AEC SETUP
AEC CALIBRATION
FLUORO SETUP
TUBE CALIBRATION
MAG:0
I/I MODES: 2
FL-RAD KV XFER:
>>
DOSE:0
MAG:0
AUTO MA/KV CURVE: 0
ABS CHANNEL: 5
DOSE:0
Action
CPI Canada Inc.
6
+
-
+
-
Ch # 740904-07