3E.2.0 ABS PICKUP INSTALLATION/WIRING
Sections 3E.1.3 and 3E.2.1 present an overview of the various ABS pickup types that may be used with
the Millenia and Indico 100 family of generators. The generator must be specifically configured to accept
each pickup type as per table 3E-2 and sections 3E.2.3, 3E.2.4, 3E.2.5 and 3E.2.6.
PLEASE NOTE THAT THE GENERATOR IS FACTORY CONFIGURED FOR ONE SPECIFIC
TYPE OF ABS PICKUP ONLY. REFER TO THE CUSTOMER PRODUCT DESCRIPTION FORM IN
CHAPTER 1 SECTION D FOR THE FACTORY CONFIGURED ABS COMPATIBILITY OF THIS
GENERATOR.
3E.2.1 Overview
The generator has been factory configured to to be compatible with one of the following ABS pickup types.
Field reconfiguration to accept other ABS pickup types, listed below, is possible if required.
PMT (photo multiplier tube).
Light sensitive optical diode. The output may be 0 to +5 VDC, 0 to -5 VDC, or +/- VDC centered at 0
VDC.
A proportional (to the brightness of the I.I) DC voltage. The output polarity may be 0 to +5 VDC or 0 to
-5 VDC.
A terminated or non terminated composite video signal.
Refer to chapter 1E for the AEC board and generator interface board location in your generator.
Page 3E-6
Figure 3E-4: ABS interface to the generator (overview)