H.V.L. EVALUATION (Cont)
ADDED ALUMINUM FILTER
0
0
0
0 (Average of three readings)
2
1
3
For 100 kVp, the HVL must be ≥ 2.7 mm Al.
X-Ray Generator Service Manual Supplement
DOSE (mR)
Table 11: HVL dose values 100 kVp
Figure 2: HVL plot 100 kVp
740917-00
Reproducibility, Linearity, & HVL Testing
RELATIVE TRANSMISSION
1.00
Rev. A
Page 11