Download Print this page

2.7 Aec Board (Universal Aec Board) - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc.
3D.2.7 AEC Board (Universal AEC Board)
The AEC board shown below is factory configured to be compatible with most makes/models of AEC
chambers (ion and solid state) on the market. This assembly also contains a low current, high voltage supply
for a photomultiplier tube (PMT). The PMT supply is located on the upper board, which also contains the
connectors to interface to the AEC pickup device(s).
AEC board input/output assignment:
Ch 1 = J1 - Table Radiographic Bucky.
Ch 2 = J2 - Vertical Wall Bucky.
Ch 3 = J3 - Spot Film Device.
Ch 4 = J4 - Aux. (Extra Bucky, Digital Acquisition, etc.).
The following potentiometers are used for AEC gain adjustment:
R1 is used for channel 1 gain adjustment.
R26 is used for channel 2 gain adjustment.
R27 is used for channel 3 gain adjustment.
R28 is used for channel 4 gain adjustment.
Refer to the end of subsection 3D.2.7 for the procedure for adjusting R10, R19, and R24.
Millenia / Indico 100 Series Service Manual
Figure 3D-10: Universal AEC board
• J7 = High voltage output for the PMT (if used)
Ch # 740904-06
Rev. D
3D
AEC Calibration