Download Print this page

Anchoring The Generator To The Floor; Leveling; Wiring To The Generator; Control Console - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc.
2.8.2
Control Console (Cont)
NOTE:
DO NOT LOCATE THE CONTROL CONSOLE WHERE X-RADIATION MAY BE PRESENT
DURING INSTALLATION OR OPERATION.
YOU MAY CHOOSE TO TEMPORARILY LOCATE THE CONSOLE NEAR TO THE GENERATOR FOR
INITIAL PROGRAMMING AND CALIBRATION. IF THIS IS SO, PLEASE COMPLETE THE FINAL
CONSOLE INSTALLATION PER THIS SECTION WHEN THE GENERATOR INSTALLATION IS
COMPLETED.
2.8.3

Anchoring The Generator To The Floor

If it is desired to anchor the generator to the floor, refer to chapter 1C. This should not be done until all
cable hookups are completed which require rear access to the generator.
2.8.4

Leveling

Adjust the leveling feet such that the generator is level and stable. This adjustment must be made for both
anchored and free-standing generator installations. As noted in 2.3.0, the leveling feet must be unscrewed
by a minimum of 1 1/2" (35 mm) to allow for proper air flow underneath the generator.
2.9.0

WIRING TO THE GENERATOR

2.9.1

Control Console

1. Connect the 15 conductor console cable from J5 at the rear of the console to J4 of the generator
interface board. Ensure that the screw locks are fully tightened to secure the connectors. Refer to
figure 2-11 for the location of J4 on the generator interface board.
2. Connect a separate ground wire, #14 AWG (2.3 mm
the console (marked CONSOLE GROUND in figure 2-3) to the ground stud located to the left of the
main fuse block on the power input board. This is marked GROUND in figure 2-7.
3. Figure 2-3 shows the designations and functions of the connectors on the rear panel of the 23 X 56
(cm) and the 31 X 42 (cm) control console.
Do not connect unapproved equipment to the rear of the console. For the 23 X 56 (cm)
console, J5 is used for the interconnect cable to the generator main cabinet, J4 is not used,
J2 is a serial port for use by an external computer, and J1 is for connection of an optional
printer.
For the 31 X 42 (cm) console, J5 is used for the interconnect cable to the generator main
cabinet, J2 is a serial port for use by an external computer, and J13 is for connection of an
external hand switch and / or foot switch. INCORRECT CONNECTIONS OR USE OF
UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE.
Indico 100 Series Service Manual
Ch # 740895-04
Rev. H
2
) or larger, from the ground stud on the rear of
2
Installation
Page 2-9

Advertisement

loading