CPI Indico 100 Series Service And Installation Manual page 195

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc.
3E.3.2 ABS Defaults (Cont)
TUBE1
ABS DEFAULT: OFF
FL TIMER MODE: 10MIN
Step
1.
2.
3.
4.
5.
6.
7.
TUBE1
LOOP GAIN
NOMINAL DOSE
DOSE 1:
DOSE 2:
Millenia / Indico 100 Series Service Manual
0%HU
FLUORO SETUP*
<<
From the FLUORO SETUP menu select >>. The above menu will display.
Select ABS DEFAULT. Toggle to select NONE, OFF, or ABS.
NONE: No default selected, ABS remains at its last setting.
OFF: Defaults to ABS OFF.
ABS: Defaults to ABS ON.
Select FL TIMER MODE. Toggle to select 5MIN or 10MIN.
5MIN: Alarms at 5.0 minutes, and stops incrementing the timer. Fluoro exposures will
continue.
10MIN: Alarms at 5.0 minutes, stops incrementing the timer at 9.6 minutes. Fluoro
exposures will be inhibited at 9.6 minutes.
Select << to return to the FLUORO SETUP menu.
Select ABS SETUP. The ABS SETUP menu will display as shown below.
Select ABS CHANNEL. This selects the hardware ABS input. This must be set to 5.
Select AUTO MA/KV CURVE. Press the + or - buttons to select the desired fluoro
mA/kV curve.
This sets how the kV and mA change during ABS operation.
0: Changes kV only (mA set manually).
1: kV linear mA.
2: mA leading kV.
3: kV leading mA #1.
4: kV leading mA #2 (reduced dose).
Refer to the graph of these curves as shown on the next page.
0%HU
*ABS SETUP*
60
80
100
100
<<
Action
AUTO MA/KV CURVE: 0
ABS CHANNEL: 5
Ch # 740904-07
Rev. D
ABS Calibration
MAG:0
DOSE:0
MAG:0
+
-
DOSE:0
3E
Page 3E-17