3B
System Interfacing
3B.6.0 TYPICAL ROOM CABLING AND INTERFACING (CONT)
Page 3B-14
Cable assembly #7
1. Must be a #10 AWG (6 mm
tubes to ground at the high tension transformer.
Cable assemblies #8
1. Pair of H.T. cables from the over table (tube-1) X-ray tube to the HT Transformer.
2. Pair of H.T. cables from the under table (tube-2) X-ray tube to the HT transformer.
Cable assembly #9.
1. Interfaces the ABS signal from the TV camera as either DC proportional or composite
video.
2. Interfaces the PMT's high voltage and signal to the generator, if used.
Refer to chapter 3E for details.
Cable assembly #10
1. These cables are usually supplied by the AEC vendor. Be sure these cables are placed
away from any electrical noise areas. When interfacing AEC cables be careful not to
cause ground loops. Grounding should only be at the AEC board.
Refer to chapter 3D for details.
Cable assembly #11
1. This cable is supplied by CPI for the remote fluoro controller interface.
Cable assembly #12
1. This cable is supplied by CPI for the control console interface.
Cable assembly #13
1. Supplied with the optional laptop computer interconnect box.
Cable assembly #14
1. AC mains cable.
Cable assembly #15
1. Cable(s) for digital imaging system.