1C
Preparing for Installation
1C.1.0 INTRODUCTION
The following items must be considered before installing your generator:
1C.2.0 GENERATOR POWER REQUIREMENTS
1C.2.1 30 kW Single Phase
Line Voltage
Line Frequency
Momentary Current
Standby Current
1C.2.2 37.5 kW Single Phase
Line Voltage
Line Frequency
Momentary Current
Standby Current
1C.2.3 30 kW Three Phase
Line Voltage
Line Frequency
Momentary Current
Standby Current
Page 1C-2
Power level of your generator.
Power line requirements.
Ground requirements.
Physical placement of the generator.
Environmental requirements for the generator.
Cable runs from the generator to all room components: tables, buckys, X-ray
tubes etc.
Rev. D
230 VAC ± 10%, 1∅.
50/60 Hz.
200 Amps.
5 Amps.
230 VAC ± 10%, 1∅.
50/60 Hz.
240 Amps.
5 Amps.
400 VAC ± 10%, 3∅ (for 400 VAC input generators).
480 VAC ± 10%, 3∅ with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC ± 10%, 3∅ (for 480 VAC input generators).
*
SEE NOTE AT THE END OF 1C.2.7
50/60 Hz.
80 Amps/phase at 400 VAC.
67 Amps/phase at 480 VAC.
5 Amps.
Indico 100 Series Service Manual
CPI Canada Inc.
Ch # 740895-02