1B
Safety
1B.6.7 mA Test Jacks Label
This label is attached to the top of the HT tank on generator models equipped with mA test jacks. A similar
warning is printed on tank lid boards using this connector, see 1B.7.8 below for details.
1B.6.8 mA test Jacks
This symbol is printed on tank lid boards with mA/mAs test jacks.
Replace the shorting jumper immediately after use. Do not attempt exposures without either
the shorting jumper in place, or an approved mA/mAs measuring device properly connected.
The warning "CAUTION: 48V MAX" printed on some tank lid boards does not apply in Millenia
generator applications.
1B.6.9 F3 (Auxiliary Power Fuseblock)
This symbol is on a label under the fuseholder for F3 (to the right of the HF power supply).
FUSE RATING: 3A 500V slow blow.
This fuse is only used if the generator is fitted with the neutral block option.
1B.7.0 SAFETY LABELS / NOTICES UNIQUE TO INDICO 100
1B.7.1 Caution HV Exposed Label
This label is attached to the removable panels on the lower (power supply) cabinet. High voltage is
present within this cabinet at all times that the mains power is switched on.
Page 1B-12
Rev. D
Millenia / Indico 100 Series Service Manual
CPI Canada Inc.
Ch # 740904-01