3C.5.3 Receptor Setup (cont)
NOTE:
IT IS RECOMMENDED THAT THE IMAGE RECEPTOR PROGRAMMING
WORKSHEET BE FILLED IN FOR EACH RECEPTOR THAT IS
PROGRAMMED. A BLANK FORM THAT SHOULD BE PHOTOCOPIED IS
LOCATED IN SECTION 3A 4.0. THIS WILL PROVIDE A RECORD OF THE
RECEPTOR SETUP FOR FUTURE REFERENCE.
If the image receptor defaults are changed, please record the original defaults in a copy of the table below:
FUNCTION
RECEPTOR
TECHNIQUE
FOCUS
FILM SCREEN
LEFT FIELD
CENTER FIELD
RIGHT FIELD
KV
MA
MS
DENSITY
NOTE:
DO NOT SWITCH OFF THE GENERATOR WHILE IN
DEFAULTS
SETUP PARAMETERS NOT TO BE SAVED. IT IS RECOMMENDED THAT THE FIRST
RECEPTOR PROGRAMMING BE COMPLETED, THE RECEPTOR SETUP MENUS BE EXITED
TO THE GEN CONFIGURATION MENU, THEN THE RECEPTOR SETUP MENU
RESELECTED TO PROGRAM THE NEXT RECEPTOR. THE ABOVE SHOULD BE REPEATED
UNTIL ALL RECEPTORS ARE PROGRAMMED. THIS WILL ENSURE THAT THE UPDATED
PARAMETERS ARE SAVED.
Use these steps to set up the receptor parameters.
Step
1.
2.
3.
4.
Millenia / Indico 100 Series Service Manual
IMAGE RECEPTOR DEFAULT SETTINGS
RECEPTOR
1
2
* MENUS 4 AND 5. DOING SO WILL CAUSE THE UPDATED RECEPTOR
From the GEN CONFIGURATION menu select RECEPTOR SETUP.
Select the desired receptor to be programmed.
Use the NEXT, and BACK buttons to navigate through the screens. As noted
earlier, the RECEPTOR SETUP [SYM] DEFAULTS menus (if enabled) are
accessed by selecting DEFAULTS in screen 3.
Select the appropriate parameter to change. Refer to the definitions on the
previous pages in this section.
RECEPTOR
RECEPTOR
3
4
Action
Ch # 740904-05
Rev. B
Programming the Generator
RECEPTOR
RECEPTOR
5
*RECEPTOR SETUP [SYM]
Page 3C-15
3C
6
BE