3C
Programming the Generator
3C.3.4 Console (cont)
Use these steps to access the CONSOLE menu.
Step
1.
2.
3.
4.
3C.4.0 APR EDITOR
The APR EDITOR enables/disables the ability of the operator to make and then save changes to APR
techniques.
Two modes of operation are available for the APR EDITOR:
ENABLED
DISABLED
GENERATORS WITH CPU BOARD 732174: When the generator is switched OFF and then
ON again, the APR editor will default to the DISABLED state. Further APR changes will
require that the APR editor be re- set to ENABLED to allow APR techniques to be changed
and then saved.
GENERATORS WITH CPU BOARD 734573: The generator stores the last APR editor setting
prior to being switched OFF. If the APR editor was previously ENABLED, APR changes
may subsequently be made and then saved in normal operating mode without the need to
manually set the APR editor to ENABLED. To disable APR technique changes, the APR
editor must be set to DISABLED.
Use these steps to set the APR EDITOR.
Step
1.
2.
Page 3C-6
From the UTILITY menu select CONSOLE.
Select the desired parameter to change. SLOW, MED, FAST KEY REPEAT and
SPEAKER VOLUME are adjusted by pressing the + or - buttons. LCD SCREEN
and APR MODE are toggled by pressing the adjacent selection button.
Select EXIT to return to the UTILITY menu.
Select EXIT again to return to the GENERATOR SETUP menu.
UTILITY
EXIT SETUP
From the GENERATOR SETUP menu select APR EDITOR.
Press the APR EDITOR button again to toggle to the desired selection.
Rev. B
Millenia / Indico 100 Series Service Manual
Action
* GENERATOR SETUP *
Allows the operator to change the default APR technique(s),
and then save the changes to memory. The APR will
subsequently default to the changed technique.
Allows temporary editing of APR technique(s), but does not
allow the changes to be saved to memory. The APR will always
default to the original technique when the generator is switched
OFF and then ON again.
Action
CPI Canada Inc.
APR EDITOR: DISABLED
GEN CONFIGURATION
DATA LINK
Ch # 740904-05