CPI Canada Inc.
3C.5.1 Tube Selection (cont)
5.
PLEASE ENSURE THAT THE SELECTED X-RAY TUBE STATOR(S) ARE COMPATIBLE
WITH THE LOW SPEED OR DUAL SPEED STARTER IN YOUR GENERATOR.
6.
3C.5.2 Generator Limits
The GENERATOR LIMITS menu allows the setting of generator limits as defined below.
MAX KW:
MAX MA:
MIN MA:
MAX MAS
EXIT
Definitions of generator limits as used in this section.
MAX KW
MAX MA
MIN MA
MAX MAS
BEFORE MAKING ANY CHANGES IN THIS SECTION, PLEASE CONSULT THE X-RAY TUBE DATA
SHEETS TO ENSURE THAT THE PROPOSED CHANGES DO NOT EXCEED THE MANUFACTURERS
RECOMMENDED LIMITS.
NOTE:
BEFORE CHANGING GENERATOR LIMITS, PLEASE FILL IN THE X-RAY
TUBE AND GENERATOR PARAMETER WORKSHEET. A BLANK FORM
THAT SHOULD BE PHOTOCOPIED IS LOCATED IN SECTION 3A 3.0. THIS
ALLOWS RECORDING OF THE DEFAULT VALUES AND THE NEW
(CHANGED) VALUES.
Use these steps to set the generator limits.
Step
1.
2.
3.
4.
Millenia / Indico 100 Series Service Manual
Select the second X-ray tube if desired by repeating the previous steps (two
tube generators only).
Select EXIT to return to the GEN CONFIGURATION menu (two tube
generators only).
* GENERATOR LIMITS*
80
1000
10
630
From the GEN CONFIGURATION menu select GENERATOR LIMITS.
Select a limit to be changed. Refer to the previous definitions.
Use the + and - buttons to change the selected values.
When finished altering the generator default values, press EXIT to return to the
GEN CONFIGURATION menu.
Sets the maximum generator kW limit.
Sets the maximum generator mA limit.
Sets the minimum generator mA limit.
Sets the maximum generator mAs limit.
Action
Ch # 740904-05
Rev. B
Programming the Generator
+
-
Page 3C-11
3C