Regular Maintenance
6.7.0

BATTERY REPLACEMENT

To replace the battery on the console CPU board or on the generator CPU board, follow the procedure below.
Refer to the figure showing the location of these batteries in chapter 2, in the section "CHECKING THE RAM
BACKUP BATTERY VOLTAGE". Refer to that section in chapter 2 for console disassembly instructions to gain
access to the console CPU board if required.
NOTE:
THE CONSOLE BATTERY SHOULD BE REPLACED WITH THE GENERATOR POWERED UP.
THIS WILL PREVENT THE CONSOLE DATA FROM BEING LOST WHEN THE BATTERIES ARE
REMOVED.
THIS IS THE ONLY EXCEPTION TO THE RULE OF NOT SERVICING THE GENERATOR WHILE
THE POWER IS ON. FAMILIARIZE YOURSELF WITH THE HIGH VOLTAGE LOCATIONS AND
HAZARDS BEFORE REPLACING THIS BATTERY.
1.
Remove the battery from the holder by gently prying under the battery at the access slot in the battery
holder using a small screwdriver. Slide the battery over the edge of the holder and remove it when it is
free.
2.
Check the voltage of the new battery prior to inserting it. This should be nominally 3.0V, do not use if it
is under 2.80 V.
3.
Wipe the replacement battery with a clean cloth, and ensure that the holder is clean and free of debris
before inserting the battery.
4.
Gently lift the spring contact on the holder and insert the replacement battery positive (+) side up
Page 6-8
Rev. D
Millenia / Indico 100 Series Service Manual
CPI Canada Inc
Ch # 740904-10