CPI Canada Inc
5.2.3
Error Messages (Cont)
ERROR CODE
E016
E017
E018
E019
E020
E021
E022
E023
E024
E025
E026
E028
E029
E030
E031
Millenia / Indico 100 Series Service Manual Ch # 740904-09
MESSAGE
TOMO BUT ERROR
NOT CALIBRATED
PREP TIMEOUT
ANODE HEAT LIMIT
THERMAL INT #1
THERMAL INT #2
DOOR INTERLOCK
COLLIMATOR ERR
CASSETTE ERROR
II SAFETY INT
SPARE INT
PREP SW CLOSED
X-RAY SW CLOSED
FLUORO SW CLOSED
REMOTE COMM ERR
PROBLEM
Tomo exposure exceeded
back up time.
Selected mA not calibrated
for selected kV.
Generator has been in prep
state too long.
Selected parameters will
cause X-ray tube to exceed
its programmed anode heat
limit.
X-ray tube # 1 too hot and
its thermal switch has
opened.
X-ray tube # 2 too hot and
its thermal switch has
opened.
Door is open.
Collimator is not ready.
Cassette is not ready.
II safety is not ready.
Spare input is not ready.
Prep input active during
power on initialization
phase.
X-ray input active during
power on initialization
phase.
Fluoro input active during
power on initialization
phase.
Communication error
detected with remote fluoro
control unit.
Rev. B
Troubleshooting
ACTION
1. Check exposure
technique
settings.
2. Increase tomo
back up time if
necessary.
Recalibrate X-ray
tube.
Reduce length of
time in prep state.
Reduce parameters
or wait for tube to
cool.
Wait for X-ray tube #
1 to cool.
Wait for X-ray tube #
2 to cool.
Close door.
Check collimator.
Check cassette.
Check II safety.
Check spare input.
Check prep switch
and input for short
circuit.
Check X-ray switch
and input for short
circuit.
Check fluoro switch
and input for short
circuit.
1. Check remote
fluoro control cable
for damage and
proper connection.
2. Turn power off
and then on to
Reset Generator.
Page 5-5
5