3C
Programming the Generator
3C.5.3 Receptor Setup
The RECEPTOR SETUP menus allow each of the image receptors to be programmed as defined in the
table following the example menu screens.
TUBE:
1
TOMO: NO
FLUORO: NO
SERIAL: NO
EXIT
RECEPTOR SYM: [sym]
FLUORO HANG: 30 SEC
RAD HANG: 0 SEC
LAST IMAGE HOLD: 40 MS
<<
SF/LF SWITCH: MAN
AEC BACKUP: FIXED
AEC BACKUP MAS: 500
AEC BACKUP MS: 3200
<<
THE DEFAULTS SELECTION IN THE MENU ABOVE IS ONLY AVAILABLE IF MEMORY IN MENU 2
WAS SET TO DEF.
RECEPTOR MENUS 4 AND 5 BELOW ARE ONLY ACCESSIBLE IF DEFAULTS IS ENABLED.
TECHNIQUE: AEC
FOCUS: SMALL
FILM SCREEN: 1
<<
KV:
75
MA:
320
MS:
50
<<
Page 3C-12
* RECEPTOR SETUP [sym] *
* RECEPTOR SETUP [sym]*
* RECEPTOR SETUP [sym]*
* RECEPTOR SETUP [sym] DEFAULTS*
* RECEPTOR SETUP [sym] DEFAULTS*
Rev. B
Millenia / Indico 100 Series Service Manual
AEC CHANNEL: 1
INTERFACE OPTS: 0
>>
MEMORY: DEF
REM TOMO BUT: 2000 MS
>>
DEFAULTS
LEFT FIELD: YES
CENTER FIELD: YES
RIGHT FIELD: YES
>>
DENSITY:
CPI Canada Inc.
+
-
+
-
+
-
0
+
-
Ch # 740904-05