5.3 Receptor Setup - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

3C
Programming the Generator
3C.5.3 Receptor Setup
The RECEPTOR SETUP menus allow each of the image receptors to be programmed as defined in the
table following the example menu screens.
TUBE:
1
TOMO: NO
FLUORO: NO
SERIAL: NO
EXIT
RECEPTOR SYM: [sym]
FLUORO HANG: 30 SEC
RAD HANG: 0 SEC
LAST IMAGE HOLD: 40 MS
<<
SF/LF SWITCH: MAN
AEC BACKUP: FIXED
AEC BACKUP MAS: 500
AEC BACKUP MS: 3200
<<
THE DEFAULTS SELECTION IN THE MENU ABOVE IS ONLY AVAILABLE IF MEMORY IN MENU 2
WAS SET TO DEF.
RECEPTOR MENUS 4 AND 5 BELOW ARE ONLY ACCESSIBLE IF DEFAULTS IS ENABLED.
TECHNIQUE: AEC
FOCUS: SMALL
FILM SCREEN: 1
<<
KV:
75
MA:
320
MS:
50
<<
Page 3C-12
* RECEPTOR SETUP [sym] *
* RECEPTOR SETUP [sym]*
* RECEPTOR SETUP [sym]*
* RECEPTOR SETUP [sym] DEFAULTS*
* RECEPTOR SETUP [sym] DEFAULTS*
Rev. B
Millenia / Indico 100 Series Service Manual
AEC CHANNEL: 1
INTERFACE OPTS: 0
>>
MEMORY: DEF
REM TOMO BUT: 2000 MS
>>
DEFAULTS
LEFT FIELD: YES
CENTER FIELD: YES
RIGHT FIELD: YES
>>
DENSITY:
CPI Canada Inc.
+
-
+
-
+
-
0
+
-
Ch # 740904-05