3D.2.1 AEC Setup Worksheet (Cont)
Note:
The example below is supplied for reference only. It does not represent an actual installation.
FUNCTION
Film/Screen
1.
2.
3.
Nominal optical density:
Grid ratio/SID:
Min - max kVp range:
± Density steps
Density dose
+%:
change
(per step)
-%:
Chamber type:
Regulatory AEC dose
requirements?
Is film processing maintained?
Assigned receptor name:
Are all cassettes similar?
Additional notes:
Additional notes:
Millenia / Indico 100 Series Service Manual
RECEPTOR 1
Lanex/reg
Lanex/reg
Lanex/med
Lanex/chest
1.2
1.1
12:1
8:1
60 - 120
65 - 140
±8
±8
12.5
12.5
6.25
6.25
Ion
Solid state
Yes
Yes
Yes
Yes
Table
Wall
Yes
Yes
Table 3D-2: Sample AEC setup worksheet
Ch # 740904-06
RECEPTOR 2
RECEPTOR 3
Lanex/reg
1.4
10:1
80 - 110
±8
12.5
6.25
Ion
Yes
Yes
SFD
Yes
Rev. D
3D
AEC Calibration
RECEPTOR 4
PMT/I.I.
10:1
70 - 120
±8
12.5
6.25
PMT
Yes
Yes
Digital
N/A
Page 3D-9

Advertisement

loading