1C.6.0 LOCATING THE GENERATOR CABINET AND CONTROLLER.
The generator cabinet is self standing and does not need to be supported. However, the installation
should meet the following requirements:
F ILE : IN _M T G .C D R
Indico 100 Series Service Manual
The floor must be flat and level.
The floor must be capable of supporting a load of approximately 250 lbs (115 Kg).
The generator installation area must be clean and free of dirt or debris.
If required, the generator may be anchored to the floor via the hold-down brackets.
See Figure 1C-2.
Sufficient room must be provided to allow access to the rear and side panels for
installation. Clearance must also be provided at both sides of the cabinet, at the front,
and at the rear of the upper cabinet to allow access for service. See Figure 1C-3 for
recommended clearances.
A cable conduit should be provided from the control console to the generator cabinet
to allow routing of the control cable if required. Allow for a 2 inch conduit. See Figure
1C-4.
16.4 (417)
TO P
V IE W
20.2 (513)
D IM E N S IO N S A R E IN IN C H E S (M M )
Figure 1C-2: Anchoring plates for securing the generator
Ch # 740895-02
1.2 (31)
Rev. D
Preparing for Installation
LE V E L LIN G
F E E T
(4 P L A C E S )
S PA C E R S
S U P P LIE D B Y
C U S TO M E R
(4 P L A C E S )
Page 1C-9