Download Print this page

6.4 Caution - Hazardous Voltages Label - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

1B
Safety
1B.6.4 Caution - Hazardous Voltages Label
This label is attached to the main access doors on the generator. Wait a minimum of five minutes after
the main power has been removed from the generator before opening the cabinet doors. This will
allow for internal capacitors to discharge to a safe level.
The upper door provides access to the HV power supply driver section, room interface board and
associated control circuits. The lower door allows access to the HT transformer, dual speed starter
chassis if fitted, line adjusting transformer if fitted, and access to the tube stator connections and thermal
switch connections. Additionally, the grounding location for the X-ray tube housing is located in this area.
Hazards associated with each area within are outlined in the relevant sections of this chapter.
Page 1B-10
CAUTION
HAZARDOUS VOLTAGES ARE
PRESENT INSIDE THE CABINET
FOR A PERIOD OF FIVE (5)
MINUTES AFTER DE-ENERGIZING
THIS EQUIPMENT.
Rev. D
Millenia / Indico 100 Series Service Manual
CPI Canada Inc.
Ch # 740904-01