3C.5.1 Tube Selection (cont)
NOTE:
BEFORE CHANGING X-RAY TUBE DEFAULT PARAMETERS, PLEASE FILL
IN THE X-RAY TUBE AND GENERATOR PARAMETER WORKSHEET. A
BLANK FORM THAT SHOULD BE PHOTOCOPIED IS LOCATED IN SECTION
3A 3.0. THIS ALLOWS RECORDING OF THE DEFAULT VALUES AND THE
NEW (CHANGED) VALUES.
The next three menus show the default tube limits. These menus appear after a tube has been selected in
the previous steps.
TUBE SPEED: DUAL
MAX SF KW LS: 21.3
MAX LF KW LS: 58.9
MAX KV: 150
RETURN
MAX SF MA:
ANODE HU WARNING: 80%
ANODE HU LIMIT: 90%
<<
SF STANDBY: 2.5A
LF STANDBY: 2.5A
SF MAX: 5.2A
LF MAX: 5.5A
<<
Definitions of tube defaults as used in this section.
TUBE SPEED
MAX SF KW LS
MAX LF KW LS
MAX KV:
MAX SF KW HS
MAX LF KW HS:
Millenia / Indico 100 Series Service Manual
*TUBE 1: RAD60
0.6/1.2
*TUBE 1: RAD60
0.6/1.2
320
*TUBE 1: RAD60
0.6/1.2
May be altered on dual speed generators. DUAL means that
the generator determines whether to use low speed or high
speed operation.
Sets the maximum small focus low speed kW limit.
Sets the maximum large focus low speed kW limit.
Sets the maximum kV allowed for that tube.
Sets the maximum small focus high speed kW limit.
Sets the maximum large focus high speed kW limit.
12º REV 1.4*
MAX SF KW HS: 35.2
MAX LF KW HS: 99.0
>>
12º REV 1.4*
>>
12º REV 1.4*
FIL BOOST: 200MS
FIL PREHEAT: 800MS
RETURN
Ch # 740904-05
Rev. B
Programming the Generator
+
-
+
-
+
-
3C
Page 3C-9