Troubleshooting
5.2.0
STATUS AND ERROR CODES
5.2.1
Operator Messages
These messages indicate the status of the generator. No action is required.
MESSAGE
INITIALIZATION
SPINNING ROTOR
X-RAY READY
X-RAY ON
5.2.2
Limit Messages
These messages indicate that an exposure has been requested that exceeds one or more limits.
MESSAGE
TUBE KV LIMIT
GEN KV LIMIT
TUBE MA LIMIT
GEN MA LIMIT
TUBE KW LIMIT
GEN KW LIMIT
TUBE MAS LIMIT
GEN MAS LIMIT
GEN MS LIMIT
CAL LIMIT
AEC DENSITY LIM
ANODE HEAT WARN
FL TIMER WARN
INVALID PARAM
HOUSE HEAT WARN
Page 5-2
Displayed during power up sequence.
Displayed when prep state is active.
Displayed when generator is ready to expose
Displayed during both a rad and fluoro exposure.
Requested kV not allowed as tube kV limit
has been reached.
Requested kV not allowed as generator kV
limit has been reached.
Requested mA not allowed as tube mA limit
has been reached.
Requested mA not allowed as generator mA
limit has been reached.
Requested parameter not allowed as tube
kW limit has been reached.
Requested parameter not allowed as
generator kW limit has been reached.
Requested mAs not allowed as tube mAs
limit has been reached.
Requested mAs not allowed as generator
mAs limit has been reached.
Requested ms not allowed as generator ms
limit has been reached.
Requested parameter not calibrated.
Requested density not programmed.
Anode has exceeded programmed warning
level.
Fluoro interval timer ≥ 5.0 mins.
Generator detected invalid parameter within
received message, message ignored.
X-ray tube housing heat has exceeded
housing warning limit.
Rev. B
Millenia / Indico 100 Series Service Manual
DESCRIPTION
PROBLEM
CPI Canada Inc
ACTION
None.
None.
None.
None.
None.
None.
None.
None.
None.
Recalibrate X-ray tube
or select a calibrated
parameter.
Select another density
or program requested
density step.
Wait for anode to cool.
Reset fluoro timer.
Select valid parameter.
Wait for housing to
cool.
Ch # 740904-09