AEC Calibration
3D.2.0 PRECALIBRATION SETUP
3D.2.1 AEC Setup Worksheet
Before continuing, it is suggested that a copy of the table below be filled in with all required information. Refer
to the example AEC setup worksheet on the next page.
FUNCTION
Film/Screen
1.
2.
3.
Nominal optical density:
Grid ratio/SID:
Min - max kVp range:
± Density steps
Density dose
+%:
change
-%:
(per step)
Chamber type:
Regulatory AEC dose
requirements?
Is film processing maintained?
Assigned receptor name:
Are all cassettes similar?
Additional notes:
Additional notes:
ALL RECEPTORS MUST HAVE THE SAME NUMBER OF DENSITY STEPS AND THE SAME DENSITY DOSE
CHANGE PER STEP (DENSITY SETTINGS ARE COMMON TO ALL FILM SCREENS AND RECEPTORS).
Page 3D-8
RECEPTOR 1
Table 3D-1: AEC setup worksheet
Rev. D
Millenia / Indico 100 Series Service Manual
RECEPTOR 2
RECEPTOR 3
CPI Canada Inc.
RECEPTOR 4
Ch # 740904-06