Reproducibility, Linearity, & HVL Testing
3.0
REPRODUCIBILITY (Cont)
3 point = 80% of maximum kV, 250 ms, mA to give 100 - 500 µR (1 - 5 µGy) dose.
2 point = 80% of maximum kV, mAs to give 100 - 500 µR (1 - 5 µGy) dose.
EXP No.
1
2
3
4
5
Calculate standard deviation (S) using formula at beginning of this section:
Calculate reproducibility =
EXAMPLE
EXP No.
1
2
3
4
5
Calculate standard deviation (s) using formula at beginning of this section:
Calculate reproducibility =
Page 6
DOSE (mR)
Calculate each of the differences
ie: DIFF
mR =
1
Repeat for each remaining mR
________
value.
S
= _________ (not to exceed 0.045)
mR
Table 4: Reproducibility
DOSE (mR)
249.0
245.0
244.0
242.0
243.0
Calculate each of the differences
ie: DIFF
mR =
1
Repeat for each remaining mR
_244.6_
value.
S
= _0.011_ (not to exceed 0.045)
mR
Table 5: Reproducibility
Rev. A
X-Ray Generator Service Manual Supplement
DIFFERENCE
= mR
- mR.
1
DIFFERENCE
4.4
0.4
0.6
2.6
1.6
= mR
- mR.
1
CPI Canada Inc
MAS = __________
DIFFERENCE
Square each difference. Then
calculate the sum of the difference
2
Sum of difference
= _________
S = __________
DIFFERENCE
19.36
0.16
0.36
6.76
2.56
Square each difference. Then
calculate the sum of the difference
2
Sum of difference
= _29.2_
S = _2.70_
2
2
.
mAs = _20_
2
2
.
740917-00