CPI Canada Inc.
3B.3.2 Outputs
• Outputs are via relay contacts, some of which may be configured to supply a dry contact closure or to
supply 24 VDC, 110 VAC, or 220 VAC upon closure.
• Note the following outputs and jumper configurations per the table below:
• To supply power to a grounded load, use TB1 pin 12 (for example) and jumper on "dry contacts". This
applies also to TB1 pin 1, TB2 pin 12, TB2 pin 1 and TB3 pin 12.
NOTE:
The Indico 100 generator is shipped from the factory with JW1 to JW5 configured for dry
contacts, and JW6 to JW8 configured for 24 VDC output on relay closure.
JUMPER CONFIGURATION (OUTPUTS):
ROOM INTERFACE BOARD
TB1 PINS 11 & 12 (BUCKY 3 SELECT)
TB1 PINS 1 & 2 (TOMO BUCKY SELECT)
TB2 PINS 11 & 12 (BUCKY 1 SELECT)
TB2 PINS 1 & 2 (BUCKY 2 SELECT)
TB3 PINS 11 & 12 (TOMO/BUCKY START)
TB3 PINS 1 & 2 (COLLIMATOR BYPASS)
TB4 PINS 11 & 12 (ROOM LIGHT)
TB4 PINS 1 & 2 TUBE 2 INDICATOR
TB4 PINS 1 & 3 TUBE 1 INDICATOR
TB5 PINS 8 & 9 (I.I. MAG 1)
TB5 PINS 5 & 6 (I.I. MAG 2)
TB5 PINS 2 & 3 (I.I. MAG 3)
TB6 PINS 1 & 2 (ALE OUTPUT)
TB6 PINS 11 & 12 (SPARE OUTPUT)
3B.3.3 Selecting Output Voltages
Five outputs (K1, K2, K3, K4 and K6 for bucky selects and bucky start) may be jumper configured for a
choice of the following voltages:
24 VDC 4 AMPS total.
110 VAC 2.5 AMPS total.
220 VAC 1.5 AMPS total.
NOTE:
2.5 AMPS IS AVAILABLE AT 110 VAC OR 1.5 AMPS IS AVAILABLE AT 220 VAC, BUT BOTH
ARE NOT AVAILABLE SIMULTANEOUSLY. TOTAL POWER CONSUMPTION MUST NOT
EXCEED 350 VA.
Indico 100 Series Service Manual
JW2 DRY CONTACTS OUTPUT
JW2 LIVE CONTACTS OUTPUT
JW1 DRY CONTACTS OUTPUT
JW1 LIVE CONTACTS OUTPUT
JW4 DRY CONTACTS OUTPUT
JW4 LIVE CONTACTS OUTPUT
JW3 DRY CONTACTS OUTPUT
JW3 LIVE CONTACTS OUTPUT
JW5 DRY CONTACTS OUTPUT
JW5 LIVE CONTACTS OUTPUT
DRY CONTACT OUTPUT ONLY
DRY CONTACT OUTPUT ONLY
JW6 PINS 1-2, 3-4 = 24VDC OUTPUT
JW6 PINS 2-3 = DRY CONTACT OUTPUT
DRY CONTACT OUTPUT ONLY
DRY CONTACT OUTPUT ONLY
DRY CONTACT OUTPUT ONLY
JW7 PINS 1-2, 3-4 = 24VDC OUTPUT
JW7 PINS 2-3 = DRY CONTACT OUTPUT
JW8 PINS 1-2, 3-4 = 24VDC OUTPUT
JW8 PINS 2-3 = DRY CONTACT OUTPUT
Ch # 740895-05
Rev. E
System Interfacing
ROOM INTERFACE BOARD
JUMPER CONFIGURATION
3B
Page 3B-7