CPI Canada Inc
5.2.3
Error Messages (Cont)
ERROR CODE
E101
E102
E103
E104
E105
Millenia / Indico 100 Series Service Manual Ch # 740904-09
MESSAGE
CAL_DATA LIMIT
CAL_MAX FIL ERR
CAL_MAN TERM
CAL_NO MA
CAL_MIN MA ERR
PROBLEM
Auto calibration has
exceeded data table length
due to an excessive number
of exposures.
Maximum filament current
for the selected focus has
been exceeded.
Operator released exposure
button during auto
calibration.
No mA feedback detected
during auto calibration.
Minimum generator mA was
exceeded at start of
calibration. This is usually
caused by too high a
filament standby current on
the primary and or
secondary filament. (Primary
is the current filament being
calibrated, secondary is the
other filament and applies
only to generators with two
independent filament
supplies).
Rev. B
Troubleshooting
ACTION
1. Check to see if
the filament
standby current is
too low.
2. Retry auto
calibration.
1. Check to see if
the maximum
filament current
limit can be
increased.
2. Retry auto
calibration.
Retry auto
Calibration.
Check power supply
Interface cables
between HF power
supply and generator
CPU board.
Reduce filament
standby current on
primary and/or
secondary filament.
Page 5-9