2.13.1 Setting tube type (cont)
5.
The example DIP switch setting shown in figure 2-10 is for the example in step 4 with an
incremental increase in boost time of 200 ms.
6.
If this is a two-tube installation, repeat steps 1 to 4 using DIP switch SW2 for the second tube.
7.
Please confirm all settings using a suitable tachometer to ensure proper anode RPM before
making any exposures.
NOTE:
FOR TUBES WHERE "LOW SPEED OPERATION ONLY" IS INDICATED, THE DUAL SPEED
STARTER MUST BE PROGRAMMED FOR LOW SPEED ONLY, AND WHERE "HIGH SPEED
OPERATION ONLY" IS INDICATED, THE DUAL SPEED STARTER MUST BE PROGRAMMED
FOR HIGH SPEED OPERATION ONLY. REFER TO THE TUBE SELECTION SECTION IN
CHAPTER 3C FOR THE PROCEDURE TO DO THIS.
NOTE THAT THE EXAMPLE DIP SWITCH SHOWN IN FIGURE 2 -10 IS REPRESENTATIVE OF
ONE STYLE OF SWITCH ONLY. DEPENDING ON MANUFACTURER, YOUR DIP SWITCH
STYLE MAY VARY. PLEASE NOTE THE ON/OFF POSITIONS CAREFULLY FOR YOUR UNIT.
Indico 100 Series Service Manual
Ch # 740895-04
Rev. H
2
Installation
Page 2-19