Download Print this page

CPI Indico 100 Series Service And Installation Manual page 241

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc
6.3.0
MAINTENANCE SCHEDULE (Cont)
Maintenance
Frequency
Every 6 Months AND
whenever a related
certifiable X-ray
component is replaced:
Every 12 months:
Every 5 years:
Millenia / Indico 100 Series Service Manual
Description of Preventative Maintenance
1.
Clean and re-grease all HV connections using vapour proof
compound.
2.
Clean the control console, remote fluoro control (if used) and main
cabinet as needed. REFER TO 6.5.0 CLEANING BEFORE
PROCEEDING
3.
Perform the X-ray tube auto calibration routine, refer to chapter 2.
4.
Verify the calibration of the generator, refer to chapter 4 of this
manual.
5.
Test the X-ray tube thermal switch circuits in the generator.
Disconnect the tube thermal switch(s) and verify the correct error
message, and that X-ray exposures are inhibited.
6.
For fan cooled Indico 100 generators in particular, remove
accumulated dust from the cooling vents. Vacuuming is
recommended.
7.
Perform any additional tests required by laws governing this
installation.
1.
Examine the following for any visible damage and replace any
damaged components:
The exterior of the control console and remote fluoro control if
used, including the membrane switch assembly.
The cable between the control console and the generator main
cabinet and between the remote fluoro control (if used) and
generator main cabinet.
The handswitch and fluoro footswitch (if used) and the cables
connecting these to the console.
2.
Open the generator cabinet and examine the unit for any visible
damage: missing or loose ground connections, oil leaks, damaged
cables etc.
3.
Ensure that there are no obstructions blocking any of the ventilation
holes or louvers on the generator cabinet.
Replace the lithium battery on the CPU board in the control console
and on the generator CPU board in the main cabinet. Refer to the
spares list in chapter 8 for the required part number. Refer to
subsection 6.7.0 for battery replacement procedure.
Ch # 740904-10
Regular Maintenance
Rev. D
6
Page 6-3

Advertisement