Tip:
If a component is not reinstalled immediately, store it in a desiccator. Doing so
prevents moisture adsorption and reduces equilibration time when the source is installed.
Warning:
corrosive materials, dispose of all waste materials according to local environmental
regulations.
5.
Fit the lenses and the spacers into the lens holder (see
the lens

4.9.4 Fitting the lenses and the spacers into the lens holder

Required materials:
• Chemical-resistant, powder-free gloves
• 2.5-mm hex wrench
To fit the lenses and the spacers:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
Notice:
powder-free gloves.
Notice:
and the spacers in the correct order and orientation. To ensure a tight fit, push each lens
and spacer down onto the one beneath it so that, when the magnetic stainless steel
lens (number 5) is fitted, the top of the lens is level with the lens holder.
1.
Slide the lenses and the PEEK spacers into the lens holder, starting with the lens identified
as "1".
To avoid spreading contamination with biologically hazardous, toxic, and
holder).
To avoid recontaminating the components, wear clean, chemical-resistant,
To ensure that the instrument operates correctly, you must replace the lenses
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Fitting the lenses and spacers into