Measure S-Parameters (Na Mode) - Keysight Technologies FieldFox Quick Reference Manual

Microwave
Table of Contents

Advertisement

Measure S-Parameters (NA Mode)

S-parameter syntax is described by the following:
S (out | in)
out = FieldFox receiver port
in = FieldFox source port
1. Press Preset then
2. Press then Mode then
following:
S11
1-port reflection measurement.
-
S21
2-port transmission measurement.
-
S12
Reverse 2-port transmission measurement.
-
Requires the full 2-port S-parameter option.
S22
Reverse 1-port reflection measurement.
-
Requires the full 2-port S-parameter option.
OR Press Trace 6 then
Multi-trace configuration. Learn more on page 8.
3. Press Freq/Dist then either
Freq Span
and
measurement.
4. Press Cal 5 to calibrate the measurement. See page 9
to learn more.
Continued on following page...
Preset
NA
then choose from the
Num of Traces
to enter a frequency range for the
6
then select a
Start
Stop
and
Center
or