Connect the Test Equipment

Connect the Logic Analyzer Pod to the Pulse Generator

1
2
ground clip
NOTE
3
4
5
Keysight U4164A Logic Analyzer Service Guide
Connect one U4203A Flying Lead Probe Set to Pods 1/2 of the U4164A module.
Connect the Pod 1 U4203A Flying Lead Probe Set's CLK lead to the pin strip of the SMA/Flying
Lead connector at the pulse generator's Channel 1 OUTPUT.
On all connection be sure to use the black ground clip (supplied with the U4203A Flying Lead
Probe Set) and orient the leads so that the black clip is connected to one of the SMA/Flying
Lead connector's ground pins!
Connect the Pod 1 U4203A Flying Lead Probe Set's CLK (NOT) lead to the SMA/Flying Lead
connector at the pulse generator's Channel 1 OUTPUT (NOT). Again, be sure to use the black
ground clip and orient the leads so that the black clip is connected to ground.
Connect the Pod 1 U4203A Flying Lead Probe Set's bits 2 and 10 to the SMA/Flying Lead test
connector's pin strip connector at the pulse generator's Channel 2 OUTPUT.
Connect the Pod 1 U4203A Flying Lead Probe Set's bits 6 and 14 to the SMA/Flying Lead test
connector's pin strip connector at the pulse generator's Channel 2 OUTPUT (NOT).
Testing U4164A Performance
3
29