Operating Instructions
To select manual output calibration for a particular channel, activate the ANALOG I/O
CONFIGURATION MENU (see Section 6.7.1), then press:
DIAG
Exit to return to
the main
PREV
sample display
DIAG AIO
< SET
Press SET> to select the analog output channel to
be configured. Then press EDIT to continue
DIAG AIO
< SET SET>
DIAG AIO
Now the analog output channels should either be automatically calibrated or they should
be set to manual calibration, which is described next.

6.7.4.3. Manual Analog Output Calibration

For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Calibration is done through the instrument software with a voltmeter connected across
the output terminals (Figure 6-2). Adjustments are made using the front panel keys by
setting the zero-point first and then the span-point (Table 6-11).
The software allows this adjustment to be made in 100, 10 or 1 count increments.

Table 6-11: Voltage Tolerances for Analog Output Calibration

Full Scale
0.1 VDC
1 VDC
5 VDC
10 VDC
Outputs configured for 0.1V full scale should always be calibrated manually
80
ANALOG I / O CONFIGURATION
NEXT
ENTR
AOUTS CALIBRATED: NO
CAL
SET>
CONC_OUT_2:5V, CAL
EDIT
CONC_OUT_2 RANGE: 5V
EDIT
SET>
Zero Tolerance
±0.0005V
±0.001V
±0.002V
±0.004V
Model 100E Instruction Manual
DIAG AIO
EXIT
< SET
SET>
DIAG AIO
EXIT
< SET
DIAG AIO
ON
EXIT
Toggles the
auto cal mode
ON/ OFF for
this analog
output channel
only.
EXIT
Span Voltage
90 mV
900 mV
4500 mV
4500 mV
NOTE
CONC_OUT_2 REC OFS: 0 mV
EDIT
CONC_OUT_2 AUTO CAL: ON
SET>
EDIT
AOUT AUTO CAL: ON
ENTR accepts the new setting
and returns to the previous menu.
EXIT ignores the new setting and
returns to the previous menu.
Span Tolerance
±0.001V
±0.001V
±0.003V
±0.006V
045150102 Rev XB1
EXIT
EXIT
ENTR EXIT