Table 7-4: Example Auto-Cal Sequence - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Calibration Procedures
The CALIBRATE attribute must always be set to OFF for analyzers used in US
EPA controlled applications that have IZS option installed.
Calibration of instruments used in US EPA related applications should only be
performed using external sources of zero air and span gas with an accuracy
traceable to EPA or NIST standards and supplied through the analyzer's sample
The following example sets Sequence2 to carry out a zero-span calibration every other
day starting at 01:00 on September 4, 2002, lasting 15 minutes. This sequence will start
0.5 hours later each day.
Mode and Attribute
Sequence
Mode
Timer Enable
Starting Date
Starting Time
Delta Days
Delta Time
Duration
Calibrate
The programmed Starting Time must be 5 minutes later than the real time
Avoid setting two or more sequences at the same time of the day. Any new
sequence which is initiated from a timer, the COM ports, or the contact
closures will override any sequence in progress. Note that two sequences with
different daily increments may eventually overlap.
If at any time an illegal entry is selected, (for example: Delta Days > 366) the
140
WARNING!
port (Section 7.2).
Table 7-4:
Example Auto-Cal Sequence
Value
2
Define Sequence #2
ZERO-SPAN
Select Zero and Span Mode
ON
Enable the timer
Sept. 4, 2002
Start after Sept 4, 2002
01:00
First Span starts at 01:00
2
Do Sequence #2 every other day
00:30
Do Sequence #2 0.5 h later each day
15.0
Operate Span valve for 15 min
ON
The instrument will re-set the slope and offset values for the SO
channel at the end of the AutoCal sequence
Programming NOTES
ENTR key will disappear from the display.
Model 100E Instruction Manual
Comment
clock.
045150102 Rev XB1
2