Figure 10-18: Power Distribution Block Diagram - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
standard IEC 320 power receptacle located on the rear panel of the instrument. From
there it is routed through the ON/OFF switch located in the lower right corner of the front
panel.
AC line power is converted stepped down and converted to DC power by two DC power
supplies. One supplies +12 VDC, for various valves and valve options, while a second
supply provides +5 VDC and ±15 VDC for logic and analog circuitry as well as the TEC
cooler. All AC and DC Voltages are distributed through the Relay Board.
A 6.75 ampere circuit breaker is built into the ON/OFF switch. In case of a wiring fault or
incorrect supply power, the circuit breaker will automatically turn off the analyzer.
Should the power circuit breaker trip correct the condition causing this
Display
Keypad
CPU
Mother
Board
Temperature
Sensors
PMT High
Voltage Supply
Pressure
Sensor
Gas Flow
Sensor
Sample/Cal
for Z/S and
IZS Valve
Options
045150102 Rev XB1
situation before turning the analyzer back on.
Chassis
PMT
Cooling
Cooling
Fan
Fan
RELAY
BOARD
UV Source
UV Source
Lamp
Lamp
Shutter
Power
Supply

Figure 10-18: Power Distribution Block Diagram

CAUTION
TEC
PMT
Control
Preamp
PCA
PS 1 (+5 VDC; ±15 VDC)
PS 2 (+12 VDC)
UV Source
Lamp
Shutter
Theory Of Operation
ON/OFF
SWITCH
KEY
AC POWER
DC POWER
PUMP
Sample
IZS Option
Chamber
Permeation
Heaters
Tube
Heater
AC POWER
ENTRANCE
189