6.10.2.4. Editing iDAS Parameters

Data channels can be edited individually from the front panel without affecting other data
channels. However, when editing a data channel, such as during adding, deleting or
editing parameters, all data for that particular channel will be lost, because the iDAS can
store only data of one format (number of parameter columns etc.) for any given channel.
In addition, an iDAS configuration can only be uploaded remotely as an entire set of
channels. Hence, remote update of the iDAS will always delete all current channels and
stored data.
To modify, add or delete a parameter, follow the instruction shown in section 6.10.2.2
then press:
YES will delete
all data in that
entire channel.
Moves the
display between
available
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
110
FROM THE DAS PASSWORD MENU
Edit Data Channel Menu
SETUP X.X
0) CONC:
PREV NEXT
INS
DEL
SETUP X.X
NAME:CONC
<SET
EDIT
PRINT
SET>
Press SET> key until...
SETUP X.X
PARAMETERS:2
<SET
SET>
PRINT
EDIT
SETUP X.X
EDIT PARAMS (DELETE DATA)
YES
NO
Edit Data Parameter Menu
SETUP X.X
0) PARAM=DETREP, MODE=INST
PREV NEXT
INS
DEL EDIT
Deletes the Parameter
currently displayed.
Model 100E Instruction Manual
ATIMER,
2,
900
PRNT
EDIT
EXIT
EXIT
EXIT
EXIT
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
NO returns to
the previous
menu and
retains all data.
Exits to the main
Data Acquisition
menu