6.6.2. Analog Output Calibration Mode

The analog outputs can be calibrated automatically or manually. In its default
mode, the instrument is configured for automatic calibration of all channels. Manual
calibration should be used for the 0.1V range or in cases where the outputs must be
closely matched to the characteristics of the recording device. Outputs configured
for automatic calibration can be calibrated as a group or individually. Calibration of
the analog outputs needs to be carried out on first startup of the analyzer
(performed in the factory as part of the configuration process) or whenever re-
calibration is required.
6.6.2.1. Automatic Analog Output Calibration
To calibrate the outputs as a group, activate the analog i/o configuration menu,
then press:
Exit at any time
to return to the
main DIAG
menu
If any of the channels have
not been calibrated this
message will read NO .
P/N 04316 Rev: B
STARTING FROM DIAGNOSTIC MENU
(see Section 6.6)
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED:
< SET SET>
CAL
DIAG AIO
AUTO CALIBRATING CONC_OUT_1
AUTO CALIBRATING CONC_OUT_2
AUTO CALIBRATING TEST_OUTPUT
DIAG AIO
AOUTS CALIBRATED:
< SET SET>
CAL
ENTR
EXIT
NO
EXIT
channel, the message
for that channel will be
YES
EXIT
Operating Instructions
If AutoCal has been
turned off for any
similar to:
NOT AUTO CAL
CONC_OUT_1
Exit to return to
the I/O
configuration
menu
85