6.9.6. COM Port Communication Modes

Each of the analyzer's serial ports can be configured to operate in a number of different
modes, which are listed in Table 6-17 and which can be combined by adding the Mode ID
numbers. For example, quiet mode, computer mode and internet-enabled mode would
carry a combined mode ID of 11, the standard configuration on the M100E COM2 port.
Note that each COM port needs to be configured independently.
MODE
ID
QUIET MODE
COMPUTER
MODE
SECURITY
MODE
ENABLE
INTERNET
HESSEN
PROTOCOL
MULTIDROP
PROTOCOL
ENABLE
MODEM
IGNORE
128
ERRORS
DISABLE
256
XON/XOFF
RS-485 MODE
1024
COMMAND
4096
PROMPT
96

Table 6-17: COMM Port Communication modes

DESCRIPTION
1
Quiet mode suppresses any feedback from the analyzer (iDAS reports, and
warning messages) to the remote device and is typically used when the port
is communicating with a computer program such as APICOM. Such feedback
is still available but a command must be issued to receive them.
2
Computer mode inhibits echoing of typed characters and is used when the
port is communicating with a computer program, such as APICOM.
4
When enabled, the serial port requires a password before it will respond.
The only command that is active is the help screen (? CR). See Section 6.4.1
for information on using the password feature.
8
Enables the use and configuration of the Ethernet interface. When enabled,
a new menu item INET will appear in the respective COM port menu.
Once this mode is enabled the INET sub menu will appear under the
COMMUNICATIONS MENU (see Menu Tree A-5 in Appendix A-1)
16
The Hessen communications protocol is used in some European countries. T-
API part number 02252 contains more information on this protocol.
32
Multidrop protocol allows a multi-instrument configuration on a single
communications channel. Multidrop requires the use of instrument IDs.
64
Enables to send a modem initialization string at power-up. Asserts certain
lines in the RS-232 port to enable the modem to communicate.
Fixes certain types of parity errors at certain Hessen protocol installations.
Disables XON/XOFF data flow control.
Configures the COM2 Port for RS-485 communication. RS-485 mode has
precedence over multidrop mode if both are enabled.
Enables a command prompt when in terminal mode.
Model 100E Instruction Manual