3.3. Initial Calibration

3.3.1. Basic Calibration Procedure

The following three-step procedure assumes that the instrument does not have any of
the available zero/span (Z/S) or IZS valve options installed. Chapter 7 contains
instructions for calibrating instruments with valve options. Chapter 8 contains directions
for performing for EPA protocol calibrations.
The initial calibration should be carried out with the analyzer's reporting range for
SINGLE range mode with a range span of 500 PPB (factory default settings for most
units). This will enable you to compare your results to the factory calibration.
STEP ONE: Set/verify the analog output reporting range of the M100E:
Press this button to set
the analyzer for SNGL
DUAL or AUTO ranges
To change the value of the
reporting range span, enter the
number by pressing the key under
each digit until the expected value
appears.
STEP TWO: Set the expected SO
The SO
concentration values
automatically default to
450.0 Conc.
To change this value to
the actual concentration of
the span gas, enter the
number by pressing the
key under each digit until
the expected value
appears.
045150102 Rev XB1
SAMPLE
< TST TST > CAL
SETUP X.X
CFG DAS
RNGE
SETUP X.X
MODE SET UNIT
SETUP X.X
0
0
SETUP X.X
0
0
span gas concentration.
2
SAMPLE*
< TST TST >
M-P CAL
< TST TST >
span
2
M-P CAL
0
0
RANGE = 500.000 PPB
SO2
=X.XXX
SETUP
PASS CLK MORE
EXIT
RANGE CONTROL MENU
EXIT
RANGE: 500.0 CONC
ENTR EXIT
5
0
0
.0
RANGE:
500.0 Conc
0
5
0
.0
ENTR EXIT
RANGE = 500.000 PPB
SO2 =X.XXX
SETUP
CAL
RANGE = 500.000 PPB
SO2 =X.XXX
ZERO
EXIT
CONC
SO2 SPAN CONC: 450.0 Conc
0
4
5
.0
ENTR EXIT
Getting Started
Press this button to select the
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT ignores the new setting and
returns to the RANGE CONTROL
MENU.
ENTR accepts the new setting and
returns to the
RANGE CONTROL MENU.
This sequence causes the
analyzer to prompt for the
expected SO
span
x
concentration.
EXIT ignores the new setting
and returns to the previous
display.
ENTR accepts the new setting
and returns to the
previous display..
41